TM Calibration Program

This topic introduces the process of automatic hand-eye calibration when a TM robot communicates with Mech-Mind Vision System through Standard Interface.

Before performing the following operations, please make sure that:
  • You have set up Standard Interface communication with the robot.

  • The robot is in the Manual mode, i.e., the green Manual light is on. If the robot is in the Auto mode, please press and hold the M/A button on the Robot Stick to switch to the Manual mode.

Load the Calibration program

  1. Click tm calibration programicon1 and select Project to open the window as shown below.

    tm calibration program1
  2. Click tm calibration programicon2, and select mm_calibaration in the pop-up window, and then click OK.

    tm calibration program2

Set the Start Point for Calibration

  1. Click Point Manager.

    tm calibration program3
  2. Click tm interface program icon3 in the pop-up window.

    tm calibration program4
  3. Click Controller.

    tm calibration program5
  4. Adjust the robot pose in the Joint or Tool tab, and click Move when finished.

This point is the first point where the camera captures the image for hand-eye calibration. Please make sure that the calibration board is at the center of the camera’s field of view.

tm calibration program6

Configure the IP Address of the IPC

  1. Click the tm interface program icon4 icon in the upper left corner of the mm_v01_init1 node.

    tm calibration program7
  2. Click the icons one by one as shown in the following image.

    tm calibration program8
    tm calibration program9
    tm calibration program10
  3. Enter the IP address of the IPC in the Add Device window and click OK.

    tm calibration program11

Run the Calibration Program

  1. In the Manual mode (the green Manual light is on), press the Run button on the Robot Stick to run the calibration program.

    tm calibration program12
  2. Make sure that the message “Entering the calibration process, please start the calibration in Mech-Vision” is displayed in the Console tab of Mech-Vision Log panel.

Complete Calibration in Mech-Vision

  1. In Mech-Vision, click Camera Calibration (Standard) on the toolbar, or select Camera  Camera Calibration  Standard from the menu bar.

  2. Follow the instructions based on different camera mounting methods to complete the configuration.

If the robot does not reach the next calibration point within 60 seconds, Mech-Vision will report a timeout error and stop the calibration process. In such cases, please select and run the calibration program on the teach pendant again, and restart the calibration process in Mech-Vision.

We Value Your Privacy

We use cookies to provide you with the best possible experience on our website. By continuing to use the site, you acknowledge that you agree to the use of cookies. If you decline, a single cookie will be used to ensure you're not tracked or remembered when you visit this website.