JAKA Calibration Program
This topic introduces the process of automatic hand-eye calibration when a JAKA robot communicates with Mech-Mind Vision System through Standard Interface.
Before proceeding, please make sure that you have set up Standard Interface communication with the robot. |
Run the Calibration Program
-
In the Programming control window, drag the two instructions below, or open the loaded mm_cali program.
-
Click Run on the toolbar.
-
Make sure that the message Entering the calibration process, please start the calibration in Mech-Vision is displayed in the Console tab of Mech-Vision Log panel.
Complete Calibration in Mech-Vision
-
In Mech-Vision, click Camera Calibration (Standard) on the toolbar, or select
from the menu bar. -
Follow the instructions based on different camera mounting methods to complete the configuration.
-
If the camera is mounted in the Eye to Hand mode, please refer to Complete Automatic Calibration in the Eye to Hand Scenario.
-
If the camera is mounted in the Eye in Hand mode, please refer to Complete Automatic Calibration in the Eye in Hand Scenario.
-
If the robot does not reach the next calibration point within 60 seconds, Mech-Vision will report a timeout error and stop the calibration process. In such cases, please select and run the calibration program on the teach pendant again, and restart the calibration process in Mech-Vision. |